Model 815 x-ray kvp meter / exposure time measurement meter manual

Model 815 x-ray kvp meter / exposure time measurement meter
Instructions
Instructions
Model 815 x-ray kvp meter / exposure time measurement meter
The Ecc Model 815 x-ray kvp meter / exposure time measurement meter measures the maximum or peak output voltage of the x-rays emitted by tungsten. Ecc Model 815 is a stable, digital instrument for measuring x-ray intensity for individuals. This instrument is not only used for x-ray acceleration voltage, but also for measuring the length of exposure time. Two functions can be checked on the instrument in the form of kv.
General operating instructions
Switch setting
In order to keep the operation simple and straightforward, the instrument itself has only two switches on the front panel.
On / off button
On / off button
The on / off button is used to switch the instrument. The switch is bright green. When the instrument is on, its light will turn to bright green. When the instrument is not in use, remember to turn off the button to save battery power.
Adjustment switch
The adjustment switch has two functions:
1. Initially the switch is turned on and the instrument switches to "High Resolution" mode.
2. After the Kvp reading is obtained, pressing this adjustment switch will provide additional information.
1. High resolution / high speed adjustment mode
When the instrument is turned on for the first time, it will prompt "Normal" or "High Speed" mode. When using the "Normal" mode, the 815 meter measures x-ray time to 100ms (0.1 second) Pressing the button once will put the instrument in "High Resolution" mode (History), and the X-ray exposure will take at least 200 ms (0.2 seconds) when the high speed mode (Hisp Speed ​​Mode) is to be used x Radiation exposure time must be at least 100 ms (1/10 seconds). When using the meter in high-resolution mode, the exposure time must be at least 200 ms (0.2 seconds)
When the instrument is set in high-resolution mode, he will keep it in high-resolution mode until it is turned off. To use it in high-speed or general-purpose mode, turn off the power and then turn it on again. Operation instructions are described in more detail in the following sections.
Second, additional information display
After the measurement, the kvp value will be displayed. Additional information will be displayed as soon as the button is pressed, and the meter will then display the following information:
● Waveform type
● Exposure time
At the same time return to the kvp mode.
The waveform type and exposure time will be displayed in 1.2 seconds. After the additional information is displayed, the display will revert to kvp mode, and the watch will prepare for the next reading. If the battery power is too low, the "Batt low" message prompt will also be displayed within 1.2s.
operating
Kvp measurement
In order to use the Model815 x-ray meter for kvp measurement, in many cases it only needs to turn on this instrument to directly measure. However, the detection target should be placed in the center of the lamp, so that the exposure value can be detected.
To ensure accurate readings, the following points should be remembered.
1. Current (mA)
When the target measured by Model 815 is in the proper position, its operating accuracy is reduced to 70mA, and the best accuracy is obtained between 10mA and 50mA.
2. Height
For half-wave x-rays in the 7 to 15 mA output range, its cone waveform should be 1/2 to 2 inches above the instrument. For x-rays at 25 to 100 mA output, place the instrument 8 to 16 inches above the instrument (1 inch = 2.540cm)
3. Calibration (collimation)
X-ray calibration should move to the center of the target by 2 ″ each time.
4. Exposure time
He has two modes that require different exposure times.
Exposure time must be at least 0.1 second (100ms) in "normal" or "high speed" mode
The instrument can measure the kvp value within a 100ms exposure time, in order to reduce wear and tear on the top and operate under radiation. Exposure time greater than 100ms is not necessary. When the instrument is used to measure the exposure time, a slightly longer exposure is allowed.
In high-resolution mode, the x-ray exposure time is at least 0.2s (200ms). The kvp value measured by the instrument between the first 200ms time reduces the wear and tear on the top, and operates under radiation. Exposure time greater than 200ms is not necessary. When the instrument is used to measure the exposure time, a long exposure time is allowed.
5. Waveform
In order to stabilize the X-ray when measuring the maximum or peak acceleration voltage of this instrument, the X-ray exposure time should be slightly delayed beyond 30ms. The instrument will display the peak value of the acceleration voltage of the x-ray exposure for the remaining time.
6. Initialize the reading
The x-ray detector will accumulate charge when it is not in use. When the initialization switch is turned on, the instrument requires a beam of radiation to work accurately.
Exposure time measurement
815 type kvp meter will also measure the kvp value when measuring the exposure time. In order to display the exposure time after measurement, press the white button, the exposure time will display the result in ms on the display. In the "normal" or "high speed" mode, the exposure time is less than 100ms can not be measured. X-ray exposure time is at least 100ms. If you need to measure the exposure time within 100ms, please call to buy 8700 digital x-ray table.
When the instrument is in high-resolution mode, the exposure time less than 200ms cannot be accurately measured, and the x-ray exposure time is at least 200ms. We recommend that you use the general mode to measure the exposure time.
More accurate exposure time is obtained at the low kvp / mA setting. Many high-energy x-ray exposure times will prolong the exposure time due to significant decay. We advocate measuring the exposure time in the range of 50 to 80kv and 25 to 50mA Measurement.
High resolution
The 815 watch is a precision measuring instrument specially designed to reduce damage to the x-ray head, but he has three situations where he can obtain accurate x-ray waveform samples in a longer time.
Low energy-when the value of kvp is 60kvp, the accurate reading can be improved by setting the exposure time of the x-ray 200ms and in high resolution mode.
Some x-ray detectors, especially old-fashioned instruments, can detect the peak value of x-ray exposure for a long period of time. In the general mode and high-speed mode, the instrument will wait 30ms before starting the measurement time. This waiting time is It is doubled in high-resolution mode, and the actual measurement time is doubled.
Some DC and multi-phase electron beams move very quickly, but he has a large overshoot. If the overshoot time is greater than 30ms in high-resolution mode, you will get an accurate reading of the true maximum kvp.
The following is a detailed description of the prompt information displayed by the instrument.
Click to enter the booth of Beijing Oriental Yuantong Technology Development Co., Ltd.
2020 National Education Equipment Cloud Exhibition
Purchasing QR code

Sweep, welcome attention

Official WeChat of Educational Equipment Procurement Network

Master the latest and most authoritative information in the education equipment industry

Copyright and Disclaimer:

â‘  The copyright of all works on this website that indicate "Source: China Education Equipment Purchasing Network" belongs to China Education Equipment Purchasing Network, and may not be reproduced, edited or used in other ways without authorization from this website. Works that have been authorized by this website should be used within the scope of authorization, and indicate "Source: China Education Equipment Purchase Network". Violators of this website will be held accountable for legal liabilities.

â‘¡ All works on this website that indicate "Source: XXX (non-this website)" are reproduced from other media. The purpose of this reprint is to transmit more information. This does not mean that this website agrees with its views and is responsible for its authenticity. Take direct responsibility and joint liability for infringement of such works. If other media, websites or individuals download and use it from this website, they must keep the "source of the manuscript" indicated on this website, and bear the legal responsibility of copyright and other.

â‘¢ If the content of the work, copyright and other issues are involved, please contact this website within two weeks from the date of publication of the work, otherwise it is deemed to waive the relevant rights.

Framing Vehicle

2 Seater Go Kart,Farm Utility Vehicles,Farm Atv Utility Vehicles,Framing Vehicle

Binzhou Daowang Power Co.,Ltd , https://www.dwutv.com